Primary technical parameters
1. Power adapter input: 100-240V~ 50/60Hz 1.5A
2. Internal power supply: DC 10.8V
3. Types of radiation: X-ray
4. Electric power:
· Maximum power: 0.21kw(70kV,3mA, 0.1s)
· Nominal electric power: 0.21kw(70kV, 3mA, 0.1s)
5. Tube voltage: tube voltage output is fixed at 70kV, error ±10%
6. Tube current: tube current output is fixed at 3mA, error ±20%
7. Loading time: the exposure loading time adjustment range
is 0.04s~1s, the grade is adjustable, and the grade is selected according to Primary technical parameters
1. Power adapter input: 100-240V~ 50/60Hz 1.5A
2. Internal power supply: DC 10.8V
3. Types of radiation: X-ray
4. Electric power:
· Maximum power: 0.21kw(70kV,3mA, 0.1s)
· Nominal electric power:0.21kw(70kV, 3mA, 0.1s)
5. Tube voltage: tube voltage output is fixed at 70kV, error ±10%
6. Tube current: tube current output is fixed at 3mA, error ±20%
7. Loading time: the exposure loading time adjustment range is 0.04s~1s, the grade is adjustable, and the grade is selected according to the R10 numerical system; with deviation ± 10%+1ms)
8. X-ray tube
a) X-ray tube model: D-045
b) Focal spot: 0.4mm
c) Target angle: 12.5°
d) Total filtration: 1.5mmAl/70 kV
e) Additional filtration: 0.5mmAl/70 kV
9. Distance from focus to skin: 20~23cm
10. Output radiation field: Φ6.0cm±0.1cm
Product specifications
Dimension: 114mm×363.8mm×245.6mm
Weight: 2.4KG
Battery :10.8V 2500mAh ×3×2
Exit field sized: Φ6.0cm±0.1cm
Reference LOADING conditions: 70kV, 3mA, 1s, loading interval 1s/60s, 59 exposures per hour.
Radiation output stated :the variation coefficient under the function less than 5%.the R10 numerical system; with deviation ±(10%+1ms)
8. X-ray tube
a) X-ray tube model: D-045
b) Focal spot: 0.4mm
c) Target angle: 12.5°
d) Total filtration: 1.5mmAl/70 kV
e) Additional filtration: 0.5mmAl/70 kV
9. Distance from focus to skin: 20~23cm
10. Output radiation field: Φ6.0cm±0.1cm
Product specifications
Dimension: 114mm×363.8mm×245.6mm
Weight: 2.4KG
Battery :10.8V 2500mAh ×3×2
Exit field sized: Φ6.0cm±0.1cm
Reference LOADING conditions: 70kV, 3mA, 1s, loading interval 1s/60s, 59 exposures per hour.
Radiation output stated: the variation coefficient under the function less than 5%.