The Original Woodpecker Battery for the O-Light Max Curing Light is a high-quality rechargeable lithium-ion battery. Designed for reliability and efficiency, it ensures optimal performance of the curing light during dental procedures.
Battery Specifications
-
Type: Rechargeable Lithium-ion Battery
-
Model: ICR 18490
-
Capacity: 1400mAh
-
Voltage: 3.6V
-
Adapter Input: 100V-240V~ 50/60Hz, 0.4A Max
-
Adapter Output: 5.0V / 1A
When to Replace the Battery
Replace the battery if:
- The device fails to power on after charging.
- The runtime after charging decreases significantly.
- The battery shows visible signs of damage or does not charge correctly.
Battery Replacement Instructions
1. Prepare the Device:
- Ensure the device is powered off.
2. Remove the Battery:
- Open the battery cover on the main unit.
- Gently remove the old battery and disconnect it.
3. Install the New Battery:
- Connect the new battery to the device.
- Place the new battery in the compartment.
- Securely close the battery cover.
Important: Only use original Woodpecker batteries to ensure safety and compatibility.
Charging Instructions
1. Setup:
- Connect the adapter to the charging port on the device.
- Plug the adapter into a power outlet.
2. Charging Indicators:
- The device's display screen will show a scrolling display during charging.
- Once charging is complete, the display will show "1111".
3. Charging Time:
- Full charge: Approximately 4 hours.
Note: Do not operate the device while it is charging.
Battery Maintenance Tips
-
Avoid Deep Discharge: Charge the battery before it is fully drained to extend its lifespan.
-
Storage: If the device is not in use for an extended period, remove the battery to prevent potential damage.
-
Monthly Recharge: If unused, recharge the battery every 6 months to maintain its functionality.
Additional Notes
- Always use the original Woodpecker adapter and battery for safety and optimal performance.
- Improper battery replacement can lead to equipment malfunction or safety hazards.
- Do not disassemble, squeeze, or shake the battery, as it may lead to circuit short or electrolyte leakage.