Intended Use: The Work Tip is used to heat, cut, soften, and pressurize gutta-percha during endodontic procedures.
Installation and Use
Installation:
Turn off the device before installing or removing the Work Tip.
Align the Work Tip's hexagonal plug with the socket in the Heating and Packing Instrument.
Install the Work Tip Protector to prevent scalding.
Operation:
Select a suitable Work Tip based on clinical needs.
Press the “ON/OFF” button to power on the device and select the desired preheating temperature.
Use the Heating button to start heating and apply gentle, continuous pressure to soften and pressurize the gutta-percha.
Avoid continuous heating for more than 4 seconds to prevent overheating and potential scalding.
Cleaning and Maintenance:
After use, clean, disinfect, and sterilize the Work Tip.
Do not polish or scratch the Work Tip. Replace if bent or worn to ensure even heating.
Here are some videos for your viewing. Enjoy!
1. How many times can the Work Tip be autoclaved?
The Work Tip can be autoclaved up to 100
times. Each cleaning, disinfection, and sterilization cycle should be documented.
2. What should I do if the Work Tip
shows visible stains after cleaning and disinfection?
If visible stains are present, you should
repeat the entire cleaning and disinfection process to ensure the Work Tip is thoroughly cleaned.
3. What temperature should not be
exceeded when cleaning the Work Tip?
The Work Tip should not be exposed to temperatures above 138°C during cleaning, disinfection, or sterilization.
4. What are the steps for post-operative
treatment of the Work Tip?
5. How should the Work Tip be prepared
before cleaning?
6. What is the recommended procedure for
cleaning the Work Tip?
7. How should the Work Tip be disinfected?
Disinfection should be performed no later
than 2 hours after cleaning. Use an automated washer-disinfector set to 93°C for 5 minutes with an A0 value greater than 3000.
8. What are the drying methods for the
Work Tip if not using automatic drying?
9. What should be done if the Work Tip
shows damage?
If the Work Tip is damaged, such as being
smashed or corroded, it must be discarded and not used further.
10. How should the Work Tip be stored
after sterilization?
Store the sterilized Work Tip in a medical
sterilization bag or clean sealing container, and keep it in a special storage cabinet. The storage time should not exceed 7 days; otherwise, reprocess before use.
Is to provide high quality equipments/supplies for all dental professionals.
!