Compatible With SATELEC® / DTE® / NSK®
The DTE ED14 ultrasonic tip is designed for the removal of broken files inside the root canal. Its slender 22mm tip enables precise and safe retrieval during clinical procedures.
For best results, position the patient’s head to keep the root canal horizontal with a downward inclination. Rotate the tip counterclockwise around the broken file to dislodge and remove it. Avoid applying excessive axial pressure to prevent pushing the file deeper into the canal.
Here are some videos for your viewing. Enjoy!
1. How often should the scaling tips be replaced?
• Scaling tips can undergo up to 300 sterilization cycles before they need to be replaced. Monitor for signs of reduced vibration or buzzing, as these may indicate a damaged tip that needs replacement.
2. How do I install the scaling tip on the Woodpecker / DTE Ultrasonic Scaler?
• Ensure the scaling tip, wrench, and handpiece have been sterilized before installation. Use the provided torque wrench to securely tighten the scaling tip onto the handpiece.
3. How do I adjust the water flow for the scaling tips?
• To adjust the water flow, use the water control knob. It’s best to start with the knob set to maximum, then adjust based on cooling and cleaning needs.
4. Why is my scaling tip’s vibration weak?
• A weak vibration can result from a loose scaling tip or a damaged tip. Tighten the scaling tip with the torque wrench. If the tip still vibrates weakly, it may need to be replaced.
5. How do I avoid damaging the scaling tip or teeth during use?
• Avoid pressing the tip too hard against the teeth and do not position the tip vertically on the tooth surface. Use a light, to-and-fro motion with minimal pressure to ensure effective and safe scaling.
6. What cleaning methods are recommended for the scaling tips?
• Cleaning should occur within 24 hours post-operation and can be done manually or with an automated cleaner. For automated cleaning, ensure it adheres to EN ISO 15883 standards, with proper temperature control and cleaning agents like distilled or deionized water.
7. How should I sterilize the scaling tips?
• Sterilization should follow a thorough cleaning and disinfection process. Tips can withstand up to 300 sterilization cycles at a maximum temperature of 138°C.